![Covidien Newport e360 Service Manual Download Page 106](http://html1.mh-extra.com/html/covidien/newport-e360/newport-e360_service-manual_2673907106.webp)
Calibration Procedures
5-10
Service Manual
Note:
To calibrate air/oxygen inlet regulators, and the safety regulator, use the diagnostic mode d3 (manual flow
adjustment).
Air Inlet Regulator
1.
Connect the air and oxygen gas supplies, and AC power to the ventilator.
2.
Enter diagnostic mode by performing the procedure in section
3.
Touch “Start Diagnostics” on the screen.
4.
Press the button below the
Trig
window. The display will blink.
5.
While the display is blinking, turn the adjustment knob to change the setting from “d0” to “d3”.
6.
Press the
Accept
button to confirm the selection.
7.
Press
Select
above the
Flow/t Insp
window.
8.
To select
Flow
, press the
Accept
button.
9.
To enable manual flow adjustment, press the
Manual Inflation
button.
10.
Press the button below the
Flow / t Inst
window. The display will blink.
11.
To obtain a flow of 3 L/min, turn the
adjustment
knob to change the setting from “0” to “3”.
12.
Press the
Accept
button to confirm the selection.
13.
Verify the value of
Air Gas Supply
is 15.5 ±0.5 psig. If not, adjust the air inlet pressure regulator:
a.
Using a 1/4” open-ended wrench, loosen the bottom lock nut (see
b.
Using a 1/4” open-ended wrench, rotate the top adjustment bolt clockwise to increase the pressure
or counterclockwise to decrease it.
c.
When the value of
Air Gas Supply
is 15.5 ±0.5 psig, secure the bottom lock nut.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......