Procedures
Service Manual
5-9
•
Green tube to XD106
•
Clear tube to XD105
40.
Using the same method as above, calibrate the zero pressure offset of the air gas supply pressure
transducer (XD102).
a.
Remove the yellow tubing from transducer XXD102.
b.
Adjust the setting of
t Insp
to “6”.
c.
Calibrate the value of
Air Gas Supply
to 0.0 ± 0.1 psig.
41.
Using the same method as above, calibrate the gain of air gas supply pressure transducer (XD102).
a.
Adjust the setting of
t Insp
to “7”.
b.
Tee in a calibrated pressure gauge and apply a pressure of 25 psig to the top port of transducer
XD102.
c.
Using the setting for
Resp Rate
, calibrate the value of
Air Gas Supply
to 25.0 ±0.1 psig.
d.
Reconnect the yellow tubing to the top port of transducer XD102.
42.
Using the same method as above, calibrate the zero pressure offset of oxygen gas supply pressure
transducer (XD103).
a.
Remove the blue tubing from transducer XD103.
b.
Adjust the setting of
t Insp
to “8”.
c.
Calibrate the value of
Oxygen Gas Supply
to 0.0 ± 0.1 psig.
43.
Using the same method as above, calibrate the gain of oxygen gas supply pressure transducer (XD103).
a.
Adjust the setting of
t Insp
to “9”.
b.
Tee in a calibrated pressure gauge and apply a pressure of 25 psig to the top port of transducer
XD103.
c.
Calibrate the value of
Oxygen Gas Supply
to 25.0 ±0.1 psig.
d.
Reconnect the blue tubing to the top port of transducer XD103.
44.
Turnoff the ventilator.
45.
Disconnect the air and O
2
gas supplies, and AC power.
5.5.2
Pneumatics Calibration
Note:
Before proceeding with the pneumatics calibration, the analog PCB in the ventilator must already be calibrated.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......