Maintenance, Overhaul, and Software Upgrade
2-30
Service Manual
17.
While the numbers are flashing, rotate the
adjustment
knob to change the altitude level as appropriate
(for example, if located at sea level, set altitude to “0 m / 0 ft”).
18.
To confirm the selection, press
Accept
.
19.
Wait 15 seconds to 20 seconds.
20.
Turn the ventilator off.
21.
To access the
Software Download
/
Diagnostics Mode
screen, press and hold the
Accept
button, and then
turn on the ventilator.
22.
Release the
Accept
button once the ventilator sounds a short beep.
23.
Touch the
Start Diagnostics
button on the screen.The
Diagnostic Data
screen will appear, as shown in
Figure 2-27.
Diagnostic Data Screen
24.
Press the
Trig Button
on the control panel.
25.
Rotate the
adjustment
knob to change the value in the display window to “d9.”
26.
Press
Accept
.
27.
Occlude the breathing circuit from the y-piece using a cap (do not use test lung).
28.
To start the exhalation flow sensor calibration (d9), press the
Manual Inflation
button.
29.
Wait for 30 seconds to 1 minute until the “TSI Air flow:” and “Exhale Flow:“ numbers start to increase
gradually and the air flow is delivered from the ventilator. The exhalation flow sensor calibration
requires 6 minutes to 10 minutes to complete (see
).
30.
When the flow delivered from the ventilator stops and ”Test Completed” appears on the screen, the
exhalation flow sensor calibration is successfully completed (see
).
31.
Turn the ventilator off.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......