Device Alert Messages
Service Manual
3-3
3.3
Device Alert Messages
defines the device alert violations and messages that may be displayed in the message
window. The first line in the message window displays the date and time of the error; the second
line displays the error message.
Monitored FiO
2
values differ
from setting
Defective oxygen sensor. Defective
analog board.
Replace the oxygen sensor. Replace analog
board.
No audible alarm
Defective alarm speaker, sound
processor board, or SBC.
Replace alarm speaker or sound processor
board, or SBC.
Press F1 to continue is displayed
on LCD during power on
Depleted battery on SBC board.
Replace battery on SBC board.
Defective SBC.
Replace SBC board.
LCD shows white screen only
Corrupted files on CF card.
Replace CF card.
Defective SBC board.
Replace SBC board.
LCD shows black screen only
Corrupted files on CF card.
Replace CF card.
Defective SBC board.
Verify voltage on the fifth solder pad from the
top, left side of connector LCD 1, is at least 2.5
volts. If not, replace SBC board.
Defective display board.
Replace display board.
Defective LCD display.
Replace LCD display.
Ventilator power resets
Defective SBC board.
Replace SBC board.
Defective power sequence board.
Replace power sequence board.
Software fails to download
Faulty USB download sticks.
Replace USB software sticks.
Table 3-1.
Troubleshooting Guide (Continued)
Problem
Potential cause
Suggested action
Table 3-2.
Device Alert Messages
Device alert violation
Violation message
Priority level
Definitions
Control communications
failure
Monitor μP Failed
High
The monitor processor does not respond to a
request from the control processor. The monitor
processor is not running.
Control CPU failure
Control CPU Failed
High
The control processor on the main PCB is bad.
Control exception failure
M Internal System
High
The control processor has detected an abnormal
operation, such as illegal instruction or division by
zero, that was generated by the control software.
Control RAM failure
Control RAM Failed
High
Random access memory that is used by the control
processor on the main PCB is damaged.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......