Component Removal and Replacement
4-4
Service Manual
4.5.2
Exhalation Flow Sensor Cable (Outside) (GR-CBL2123P)
To replace the exhalation flow sensor cable, perform the following steps:
1.
Remove the exhalation valve and exhalation flow sensor by following the steps in section
(EXH2105A) and Flow Sensor (FLS2101P)
2.
Remove the exhalation flow sensor cable (outside) (GR-CBL2123P) from the exhalation manifold by
pulling the collet of the cable connector.
Note:
The lock mechanism of the connector can be easily released by pulling the collet of the connector. Do NOT
disconnect the connector by pulling the wire.
3.
Reinstall the cable to the exhalation manifold by aligning the key on the connector with the notch of the
female connector on the exhalation manifold.
4.6
Top Cover (CVR2101M)
Figure 4-2.
Top Cover
1
Screws (three of six securing cover to ventilator)
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......