Maintenance, Overhaul, and Software Upgrade
2-22
Service Manual
Figure 2-18.
Verifying the Time and Date
2.9
Biannual Maintenance
2.9.1
Internal Battery
Replace the internal battery at least every 2 years, or as needed to meet performance
requirements.
To remove and install the internal battery, follow the procedure in section
, “
WARNING:
Observe correct polarity when reconnecting the battery connectors.
Caution:
To avoid discharging the battery voltage, do not allow any metal object (including tools) to touch the
battery connectors.
2.9.2
Oxygen Sensor
Replace the oxygen sensor at least every 2 years, or as needed to meet performance
requirements.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......