Procedures
Service Manual
5-13
5.
Adjust
Flow
to 0 L/min once the calibration is complete.
6.
To disable the manual flow adjustment, press the
Manual Inflation
button.
Air PSOL Valve: d1
Caution:
Do not attempt to perform air PSOL valve calibration unless instructed by the Technical Service
Department or if one of the following conditions occurred:
•
Replacement or readjustment of air inlet regulator
•
Replacement of air PSOL valve
•
Replacement of analog PCB
•
Replacement of main PCB
Caution:
Do not attempt to perform air PSOL valve calibration unless the air gas supply is regulated to 50 ±2 psig
and capable of supporting a minimum of 180 L/min of constant flow.
1.
Enter diagnostics mode by following the procedure in section
2.
Touch “Start Diagnostics” on the screen.
3.
Press the button below the
Trig
window.The display will blink.
4.
While the display is blinking, turn the adjustment knob to change the setting from “d0” to “d1”.
5.
Press the
Accept
button to confirm the selection.
6.
Disconnect the patient circuit from the ventilator.
7.
To initiate air PSOL valve calibration, press the
Manual Inflation
button.
•
The ventilator will go through a calibration maneuver that slowly increases and then decreases flow
twice.
•
The value of
TSI Air Flow
displayed on the screen will slowly change as the delivered flow changes.
8.
Wait several minutes for the calibration to complete, and the message “Test Completed” to be displayed.
Note:
If the message “Calibration Failed: Valve Leaks more than 50 cc” is displayed after the d1 is complete, the PSOL
valve needs to be replaced.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......