Calibration Procedures
5-8
Service Manual
Note:
If pressure is only applied to transducer XD105 during this calibration, a device alert will occur when the
pressure to XD105 is 20 cmH
2
O higher than the pressure to XD106. A device alert can be only cleared by
turning off and on the ventilator.
19.
Adjust the pressure syringe so that 110 cmH
2
O is displayed on the calibrated pressure gauge.
20.
Adjust
t Insp
to “1” (Drive Pressure).
21.
Press the
Manual Inflation
button. The message “Start Calibration” is displayed.
22.
Press the button below the
Resp Rate
window. The display will blink.
23.
While the display is blinking, turn the adjustment knob to adjust the value.
24.
Press the
Accept
button to confirm the selection. The value of Drive Pressure displayed on the screen
changes.
25.
Repeat steps 22 and 23 to obtain a Drive Pressure reading of 110.00 ±1.0 cmH
2
O.
26.
Press the
Manual Inflation
button to confirm the selection. The message “Test Completed” is displayed.
27.
Reset the respiratory rate to 128.
28.
Set
t Insp
to “3” (Machine 2).
29.
Press the
Manual Inflation
button. The message “Start Calibration” is displayed.
30.
Repeat steps 22 and 23 to obtain a Drive Pressure reading of 110.00 ±1.0 cmH
2
O.
31.
Press the
Manual Inflation
button to confirm the selection. The message “Test Completed” is displayed.
32.
Reset the respiratory rate to 128.
33.
Set
t Insp
to “5” (Machine 1).
34.
Press the
Manual Inflation
button. The message “Start Calibration” is displayed.
35.
Repeat steps 22 and 23 to obtain a Drive Pressure reading of 110.00 ±1.0 cmH
2
O.
36.
Press the
Manual Inflation
button to confirm the selection. The message “Test Completed” is displayed.
37.
Reset the respiratory rate to 128.
38.
Disconnect the low pressure calibration tool (TOL2108A) from the analog aboard.
39.
Reconnect the following tubes from inside the ventilator:
•
Orange tube to XD107
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......