Troubleshooting
3-4
Service Manual
.
Control ROM failure
Control ROM Failed
High
Read-only memory that stores the code of the
control processor has an incorrect checksum.
Control task continuity failure
Control Tasks Failed
High
Software tasks of the control processor have
operated out of sequence.
Dual RAM failure
Dual RAM Failed
High
Random access memory that is shared between
the control and monitor processors is damaged.
Monitor communications
failure
Control uP Failed
High
The control processor does not respond to a
request from the monitor processor. The control
processor is not running.
Monitor CPU failure
Monitor CPU Failed
High
The monitor processor on the main PCB is bad.
Monitor exception failure
C Internal System
High
The monitor processor has detected an abnormal
operation, such as illegal instruction or division by
zero, that was generated by the monitor software.
Monitor RAM failure
Monitor RAM Failed
High
Random access memory that is used by the
monitor processor on the main PCB is damaged.
Monitor ROM failure
Monitor ROM Failed
High
Read-only memory that stores the code of the
monitor processor has an incorrect checksum.
Monitor task continuity failure
Mon Task Failed
High
Software tasks of the monitor processor have
operated out of sequence.
Power failure
Power Failure
High
DC power out of tolerance. Check +12 VDC, –12
VDC, and +5 VDC. (The ventilator may have been
powered by internal battery until it was depleted
and a device alert resulted.)
Pressure sensor error
Pressure Sensor Error
High
Loss of pressure to machine 1 or machine 2
pressure transducers. (Can be caused by
disconnected tube, failure of rezero solenoid, or a
faulty pressure transducer.)
Table 3-2.
Device Alert Messages (Continued)
Device alert violation
Violation message
Priority level
Definitions
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......