Theory of Operations
A-2
Service Manual
The exhalation system is heated to prevent moist exhaled gas from condensing in the exhalation
pathway. A bacteria filter should be used at the From Patient port to prevent contaminants in the
exhaled gas from entering the exhalation system and contaminating the exhalation valve and flow
sensor. Another filter should be placed at the To Patient port to prevent contamination of the
inspiratory manifold when the emergency relief valve opens (in the case of a Device Alert alarm,
Both Air/O
2
Supply Loss alarm, or Sustained High Baseline Pressure alarm that is accompanied by a
blockage of the exhalation valve).
The heated exhalation system features an active exhalation valve with a low exhaled flow resistance
for rapid return of circuit pressure to baseline and decreased potential for auto-PEEP.
A.2
Mandatory Breath Types
The ventilator offers these breath types:
•
Volume Control
•
Pressure Control
•
Biphasic Pressure Release Ventilation*
•
Volume Target Pressure Control*
* Not available on the S model
A.2.1
Volume Control
Volume Control ventilation provides time-cycled, volume-limited mandatory breaths.
Volume
and
Flow
(or inspiratory time) are user-set parameters and pressure is allowed to vary. The user has the
option to select between two flow wave patterns for breath delivery in volume ventilation. The flow
wave pattern function is accessed via the
Advanced
data set.
A square flow wave pattern delivers the set flow constantly until the set tidal volume is delivered.
Reference
Table A-1.
Control Selections for Mandatory Breaths
Mandatory breath type
Control panel selection
Advanced selection
open exhalation
Advanced selection
volume target
Volume Control
Volume Control
Not available
Off
Pressure Control
Pressure Control
Off
Off
Volume Target Pressure
Control (VTPC)
Volume Control or Pressure
Control
Off
On
Biphasic Pressure Release
(BPRV)
Pressure Control
On
Off
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......