4.24
Battery (BAT2100A)
Figure 4-27.
Battery (BAT2100A)
To replace the battery, refer to
and perform the following steps:
1.
Remove the top cover by performing the steps in section
2.
Remove the battery wires from the positive and negative terminals.
3.
Remove the screw securing the battery bracket to the ventilator.
4.
Lift the bracket off of the ventilator.
5.
Remove the battery by lifting it up and out.
6.
Reinstall the battery by reversing steps 2 through 5.