2-9
Flange shape
The flanges used should be such that the area of contact with the gasket is maximized,
as shown in Fig. 2-6.
Acceptable
X Unacceptable
Flange
Welding
Welding
Pipe
(The liquid could
leak because of
the small area
of contact with
the gasket.)
Fig. 2-6. Flange shape
Note) Before installing the detector be sure to flush out any foreign matter that may be present
in inside of the detector. Residual foreign matter could cause output fluctuations.
Do not touch the electrodes by your hand or waste cloth with oil. It could cause output
fluctuations.
Install the detector in accordance with the flow direction mark on the detector with the
direction of the liquid flow. Misalignment could result in a negative output.