5-33
5-5-10. Entering ENGINEERING MODE and MAINTENANCE MODE
Introduction
This section describes how to enter ENGINEERING MODE, in which setup parameters for
the electromagnetic flowmeter are to be configured, and MAINTENANCE MODE, in which
calibration and check are to be carried out.
Note) The mode selection screen may not appear, depending on the settings of write protect.
Operate the write protect switch on the main board, and then select one from levels 1,
2 and 3 to display the screen for selecting ENGINEERING MODE only. Select write
protect level 0 to display the screen for selecting both ENGINEERING MODE and
MAINTENANCE MODE. See “5-3-2. Display of write protect level”.
The procedure for entering ENGINEERING MODE is shown below.
Step
Procedure
Screen
1
The screen shown right is a screen display example in
MEASURING MODE (measurement state).
Press the MODE key.
20.0 %
01.94 m
3
/h
WP0
00069401
2
The OPERATOR'S MODE screen appears for approx.
two seconds, and then the damping setting screen
appears.
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
3
Press the
key twice to display the screen as shown.
20.0 %
* MODE ENTER
ENGINEERING
4
Press the
key once to move the cursor to the
position shown on the screen.
20.0 %
* MODE ENTER
ENGINEERING
5
Press the
key.
With the display is changed, ENGINEERING MODE
is active.
The screen appears in approx. two seconds.
20.0 %
# ENGINEERING
MODE
20.0 %
# ID SET
XXXXXXXX
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......