5-46
Set pulse scale in accordance with the following procedure:
Step
Procedure
Screen
1
Enter ENGINEERING MODE in accordance with the
entry into ENGINEERING MODE (see section 5-5-
10. on page 5-33). Then press the
or
key to
display the screen at right.
12.3 %
# PLS 27.780 Hz
100.00 l/p
SCL
2
Press the
key to move the cursor to the position
under a desired value to be set or changed.
12.3 %
# PLS 27.780 Hz
100.00 l/p
SCL
3
Using the
or
key, change the value to a desired
pulse scale to be set.
12.3 %
# PLS 13.890 Hz
200.00 l/p
SCL
4
Press the
key to move the cursor to the position
under #.
Press the MODE key to return to the MEASURING
MDOE and to save data.
12.3 %
# PLS 13.890 Hz
200.00 l/p
SCL
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new value before
the system resets it to the previously saved value.