7-38
Tag setting
To set the tag, proceed as follow:
1. From online menu,
Select:
1. Device setup
3. Basic setup
1. Tag
Fig. 7-92. will then be displayed.
Fig. 7-92.
2. Once the display as shown in Fig. 7-92. appears, use the arrow key up, down, right or left
and numerical keys to enter the number. After entering, press F4 (ENTER). If F3 (ESC) is
pressed here, the selection will be canceled and the display will return to the Basic setup
menu.
3. After returning to the Basic setup menu, press F2 (SEND). Changed setting will be
transmitted to the converter.
Shipping data recovery
Follow the procedure described below to execute the shipping data recovery. Please make sure
that when it is executed, the internal data will return to the factory setting.
1. From online menu,
Select:
1. Device setup
2. Diag/Service
3. Calibration
6. Shipping RCVR
Fig. 7-93. will then be displayed.
Fig. 7-93.
2. When executing shipping data recovery, select “YES” and press F4 (ENTER). Shipping
data recovery will execute, and the display will return to Fig. 7-93. Press F3(ABORT) to
return to the previous menu.