7-3
7-1-3. Verifying communication
After the HART Communicator has been properly interconnected, turn the device's power on.
For the external power supply model, turn on the external power supply before turning the
device power on.
Once the setting and wiring connections are correct, the HART Communicator's display
shows an online menu as shown below and a HART mark will flicker in the upper right hand
corner of the display.
Fig. 7-2. Online menu
If the display is not as shown in Fig. 7-2. but as shown in Fig. 7-3. below, no communications
are being made. Recheck the HART Communicator connections and the setting of converter.
(The setting of the converter is described page 7-2 .)
HART Communicator
No Device Found
Press OK
SAVE
OK
Fig. 7-3. Communication not available
7-1-4. Cautions
CAUTION
Do not remove the HART Communicator cable from the converter while executing
communication. If the cable is disconnected during data setting transmission there will be
no data transfer to the converter.