5-26
Step
Procedure
Screen
9
(Continued)
(Branch 4)
When the result is (4) in Step 8
Check that the empty or scale on electrode status is
not detected when SENSITIVITY MID has been set.
(Result)
If the empty or scale on electrode status is not
detected, use that SENSITIVITY MID setting
without change.
If the empty or scale on electrode status is detected
and the screen shown on the right appears, this
function is not available in that environment. Set
the function to OFF.
0.0 %
EMPTY OR SCALE
ON ELECTRODE
(Branch 5)
When the result is (5) in Step 8
Check that the empty or scale on electrode status is
not detected when SENSITIVITY HIGH has been
set.
(Result)
If the empty or scale on electrode status is not
detected, use that SENSITIVITY HIGH setting
without change.
If the empty or scale on electrode status is detected
and the screen shown on the right appears, this
function is not available in that environment. Set
the function to OFF.
0.0 %
EMPTY OR SCALE
ON ELECTRODE