7-4
7-2. Setting and calibrating devices using the HART Communicator
The HART Communicator enables the user to set a two wired magnetic flowmeter device as
well as allowing them to adjust and check the output of the device and to inspect the device.
The following values can be set using the HART Communicator:
Flow unit
Low flow cut
Range
Drop out
Specific gravity
Pulse scale unit
Damping time constant
Pulse scale
Display selection
Pulse width
Function selection
Totalized value of integral counter display
Correction coefficient
Reset value of integral counter
Communication method
Integral counter reset
Detector diameter
High alarm value setting
Detector constant
Low alarm value setting
Auto spike cut
Setting output status setting
Average processing selection
Burnout (Analog output) setting
Average processing time
Burnout (Pulse output) setting
Electrode status diagnostic function
Also, the following calibrations and inspection can be made:
Zero point adjustment
Current output calibration at 4 mA and 20 mA
Gain adjustment
Pulse output adjustment
Excitation current output adjustment
Analog output check
Pulse output check
Contact output check
Converter status check
Tag setting
Shipping data recovery
Equipment information check
For a detailed list of all the menus, see the HART Communicator's menu table for the two
wired magnetic flowmeter at the back of this manual.
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......