6-72
Checking pulse output [PULSE OUTPUT] function
Pulse output can be checked by fixing the pulse output from the converter main body by using
SFC.
Use the following procedure to check the pulse output.
Step
Key
Procedure
SFC screen
1
CLR
( No )
Confirm that the SFC is set to “READY”.
If it is not, press the
CLR
( No )
key to set it
to “READY”.
SFCM00006009D
2
B
CONF
H
NEXT
or
L
PREV
Press the
B
CONF
key to access the
CONFIG functions. Press the
H
NEXT
key or
L
PREV
key to display the screen
as shown on the right.
SFCM00006136D
SFCM00006020D
3
ENTER
( Yes )
NON-VOL
H
NEXT
or
L
PREV
Press the
ENTER
( Yes )
NON-VOL
key to access the
TOTALIZER functions. Press the
H
NEXT
key or
L
PREV
key to display
the screen as shown on the right.
SFCM00006137D
4
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key to display the
screen as shown on the right.
SFCM00006138D
5
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key, again.
The actual pulse output will be displayed
as shown on the right.
SFCM00006139D
6
Enter the output value to be checked
into the screen as shown on the right
using numeric key.
SFCM00006140D
Summary of Contents for MagneW Two-wire PLUS+ MTG15A