6-60
Select pulse output / contact output [DIGITAL I/O] function
Two wired magnetic flowmeter can be configured for pulse output or contact output other
than the analog current output (4-20 mA).
Use the following procedure to configure the pulse output and contact output.
Step
Key
Procedure
SFC screen
1
CLR
( No )
Confirm that the SFC is set to “READY”.
If it is not, press the
CLR
( No )
key to set it
to “READY”.
SFCM00006009D
2
B
CONF
H
NEXT
or
L
PREV
Press the
B
CONF
key to access the
CONFIG functions. Press the
H
NEXT
key or
L
PREV
key to display the screen
as shown on the right.
SFCM00006100D
3
ENTER
( Yes )
NON-VOL
H
NEXT
or
L
PREV
Press the
ENTER
( Yes )
NON-VOL
key to access the
DIGITAL I/O function, press the
H
NEXT
key or
L
PREV
key to display
the screen as shown on the right. Present
setting function appears.
DO NOT USED..... Pulse output
HI-LO ALARM..... Contact output
* Following function will appear on the
screen but are not available for two
wired magnetic flowmeter.
ALARM
DIAG
EMPTY
H1-L1/H2-L2 ALM
SFCM00006101D
or
SFCM00006102D
4
MENU I
ITEM
DE CONF.
Press the
MENU I
ITEM
DE CONF.
key and specify the
function to be set.
SFCM00006102D
5
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key.
The changed setting is saved to the SFC.
SFCM00006103D
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......