7-25
7-3. Calibrating and Inspecting the device by HART Communicator and other
functions
7-3-1. Device adjustment
Analog current output adjustment
To adjust analog output (4 mA and 20 mA) adjustment, proceed as follows:
1. From online menu,
Select:
1. Device setup
2. Diag/Service
3. Calibration
1. D/O trim
Fig. 7-43. will then be displayed.
If the control system is not affected by the
forcible change of the current signal, press
F4 (OK). By pressing F3 (ABORT), the
procedure is canceled.
OK
EXIT
ENTER
ESC
ABORT
DEL
SAVE HOME
HELP
MAGNEW2W: ** ******
Setting fld dev
Output to 4 mA
Fig. 7-43.
2. Once the display as shown in Fig. 7-44.
appears, move the arrow key up or down
to select a current output to adjust, and
then press F4 (ENTER). This time, select 4
mA.
OK
EXIT
ENTER
ESC
ABORT
DEL
SAVE HOME
HELP
MAGNEW2W: ** ******
Choose DAC trim func
1 Set
2 End
Fig. 7-44.
3. Fig. 7-45. will then appears. Connect the
device to measure the current output, and
then press F4 (OK).
OK
EXIT
ENTER
ESC
ABORT
DEL
SAVE HOME
HELP
MAGNEW2W: ** ******
Enter meter value
4.05
Fig. 7-45.