7-36
Contact output check
To output contact fixed value, proceed as follows:
1. From online menu,
Select:
1. Device setup
2. Diag/Service
2. Loop test
4. Out put check Do
Fig. 7-87. will then be displayed.
If the control system is not affected by the
forcible change of the current signal, press
F4 (OK). By pressing F3 (ABORT), the
procedure is canceled.
Fig. 7-87.
2. Fig. 7-88. will then appear. Choose
START to start fixed value of contact
output, and then press F4 (ENTER)
Fig. 7-88.
3. Fig. 7-89. will appear. Move the arrow key
up or down to select an OPEN or CLOSE.
After making a selection, press F4
(ENTER). Selected contact will be output.
Press F3 (ABORT) to return to the
previous menu.
Fig. 7-89.