6-2-3. Setting using SFC communication (2) - setting using CONFIG functions
Selecting unit system and setting specific gravity [UNIT KEY] function
It is possible to select unit system (Mass flow rate and volume flow rate) which is set by two
wired magnetic flowmeter converter, and set the specific value (in case of selecting mass flow
rate for system units). Use the following procedure to select unit system and to set specific
gravity.
Step
Key
Procedure
SFC screen
1
CLR
( No )
Confirm that the SFC is set to “READY”.
If it is not, press the
CLR
( No )
key to set it
to “READY”.
SFCM00006009D
2
B
CONF
Press the
B
CONF
key to access the
CONFIG functions. Confirm that the
screen on the right appears.
SFCM00006079D
SFCM00006080D
3
ENTER
( Yes )
NON-VOL
MENU I
ITEM
DE CONF.
Press the
ENTER
( Yes )
NON-VOL
key. The screen, which
system of units can be selected, appears
as shown to the right. Mass flow rate =
MASS FLOW Volume flow rate =
VOLUME FLOW
Press the
MENU I
ITEM
DE CONF.
key, and select MASS
FLOW or VOLUME FLOW.
SFCM00006081D
4
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key. The screen as
shown on the right appears and the
changed setting is saved to the SFC.
SFCM00006082D
5
H
NEXT
or
L
PREV
Only when MASS FLOW is selected, a
screen for setting the specific gravity
appears.
Press the
H
NEXT
key or
L
PREV
key to
show this screen. Press numeric keys to
set specific gravity. Available range of
specific gravity is from
0.1000 to 5.9999.
SFCM00006083D
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......