6-80
Setting counter reset function [RESET TOTALZE] function
Resetting the totalized value and setting reset value are possible using the SFC. When resetting
the totalized value, two wired magnetic flowmeter internal counter will be reset and start
resetting from the reset value which has been set. You can enter the reset value maximum of
10 figures to SFC, however for two wired magnetic flowmeter only 8 figures are effective. Use
the following procedure to reset the totalized value and to set the reset value.
Step
Key
Procedure
SFC screen
1
CLR
( No )
Confirm that the SFC is set to “READY”.
If it is not, press the
CLR
( No )
key to set it
to “READY”.
SFCM00006009D
2
B
CONF
H
NEXT
or
L
PREV
Press the
B
CONF
key to access the
CONFIG functions. Press the
H
NEXT
key or
L
PREV
key to display the screen
as shown on the right.
SFCM00006136D
SFCM00006020D
3
ENTER
( Yes )
NON-VOL
H
NEXT
or
L
PREV
Press the
ENTER
( Yes )
NON-VOL
key to access the
TOTALIZER functions. Press the
H
NEXT
key or
L
PREV
key to display
the screen as shown on the right.
SFCM00006155D
4
ENTER
( Yes )
NON-VOL
Press the
ENTER
( Yes )
NON-VOL
key.
Displays the actual reset value.
SFCM00006156D
5
Enter the reset value to be set pressing
the numeric keys.
SFCM00006157D
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......