![Agilent Technologies Agilent 4396B Service Manual Download Page 317](http://html.mh-extra.com/html/agilent-technologies/agilent-4396b/agilent-4396b_service-manual_2867991317.webp)
211
GND
LEVEL
OUT
OF
SPEC
The
voltage
of
the
GND
(Ground)
at
the
DC
bus
node
26
is
out
of
its
limits
.
This
message
is
displayed
when
an
internal
test
4:
A2
POST
REGULA
TOR
fails
.
Troubleshoot
the
power
supply
functional
group
in
accordance
with
the
P
ower
Supply
Troubleshooting
chapter
.
205
HP-HIL
CHIP
TEST
F
AILED
The
A1
CPU's
HP-HIL
control
chip
does
not
work
properly
.
This
message
is
displayed
when
an
internal
test
1:
A1
CPU
fails
.
Replace
the
A1
CPU
with
a
new
one
.
See
the
Digital
Control
Troubleshooting
chapter
.
204
GPIB
CHIP
TEST
F
AILED
The
A1
CPU's
GPIB
chip
does
not
work
properly
.
This
message
is
displayed
when
an
internal
test
1:
A1
CPU
fails
.
Replace
the
A1
CPU
with
a
new
one
.
See
the
Digital
Control
Troubleshooting
chapter
.
255
IF
GAIN
SWITCHING
UNC.
OUT
OF
SPEC
This
message
is
displayed
when
an
external
test
37:
IF
GAIN
fails
.
Troubleshoot
the
analyzer
in
accordance
with
the
Isolate
F
aulty
Group
Troubleshooting
chapter
.
229
INPUT
A
TTENU
A
TOR
A
CCURA
CY
OUT
OF
SPEC
This
message
is
displayed
when
an
external
test
21:
INPUT
A
TTENU
A
TOR
fails
.
Troubleshoot
the
analyzer
in
accordance
with
the
Isolate
F
aulty
Group
Troubleshooting
chapter
.
202
KEY
CHIP
TEST
F
AILED
The
A1
CPU's
front
keyboard
control
chip
does
not
work
properly
.
This
message
is
displayed
when
an
internal
test
1:
A1
CPU
fails
.
Replace
the
A1
CPU
with
a
new
one
.
See
the
Digital
Control
Troubleshooting
chapter
.
228
OUTPUT
A
TTENU
A
TOR
A
CCURA
CY
OUT
OF
SPEC
This
message
is
displayed
when
an
external
test
20:
OUTPUT
A
TTENU
A
TOR
fails
.
Troubleshoot
the
analyzer
in
accordance
with
the
Isolate
F
aulty
Group
Troubleshooting
chapter
.
48
PHASE
LOCK
LOOP
UNLOCKED
A
phase
lock
loop
(PLL)
circuits
within
the
analyzer
does
not
work
properly
.
Troubleshoot
the
analyzer
in
accordance
with
the
Digital
Control
Troubleshooting
chapter
.
When
a
Svc
annotation
is
displayed
(Service
Modes
are
activated),
this
error
message
does
not
appear
even
if
a
PLL
circuit
is
not
working.
210
POST
REGULA
TOR
OUTPUT
V
OL
T
A
GE
OUT
OF
SPEC
A
power
supply
voltage
of
the
A2
post-regulator
is
out
of
its
limits
.
This
message
is
displayed
when
an
internal
test
4:
A2
POST
REGULA
TOR
fails
.
Troubleshoot
the
power
supply
functional
group
in
accordance
with
the
P
ower
Supply
Troubleshooting
chapter
.
49
PO
WER
F
AILED
ON|-
P
ower
failure
occurs
on
the
power
lines
listed
in
the
message
.
One
or
some
of
+15
V
,
+5
V
,
05
V
,
015
V
,
and
P
ostRegHot,
follow
the
message
.
Troubleshoot
the
power
supply
functional
group
in
accordance
with
the
P
ower
Supply
Troubleshooting
chapter
.
Messages-5
Summary of Contents for Agilent 4396B
Page 10: ......
Page 32: ......
Page 43: ...Figure 2 7 CAL OUT Level Adjustment Location Adjustments and Correction Constants 2 11 ...
Page 46: ...Figure 2 10 Comb Generator Output 2 14 Adjustments and Correction Constants ...
Page 62: ...Figure 2 26 Final Gain Adjustment Location 2 30 Adjustments and Correction Constants ...
Page 76: ...Figure 3 1 Troubleshooting Organization 3 2 T roubleshooting ...
Page 84: ......
Page 90: ...Figure 5 1 Power Supply Lines Simpli ed Block Diagram 5 2 Power Supply T roubleshooting ...
Page 107: ...Figure 5 12 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 19 ...
Page 108: ...Figure 5 13 Power Supply Block Diagram 2 5 20 Power Supply T roubleshooting ...
Page 109: ...Figure 5 14 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 21 ...
Page 110: ......
Page 112: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 124: ......
Page 126: ...Figure 7 1 Source Group Block Diagram 7 2 Source Group T roubleshooting ...
Page 160: ...Figure 8 1 Receiver Group Simpli ed Block Diagram 8 2 Receiver Group T roubleshooting ...
Page 168: ......
Page 184: ...Figure 10 6 External Test Setup 1 Figure 10 7 External Test Setup 2 10 10 Service Key Menus ...
Page 185: ...Figure 10 8 External Test Setup 3 Figure 10 9 External Test Setup 4 Service Key Menus 10 11 ...
Page 226: ...Figure 11 3 Power Supply Functional Group Simpli ed Block Diagram 11 6 Theory of Operation ...
Page 231: ...Figure 11 5 Digital Control Group Simpli ed Block Diagram Theory of Operation 11 11 ...
Page 235: ...Figure 11 6 Source Simpli ed Block Diagram Theory of Operation 11 15 ...
Page 244: ...Figure 11 7 Receiver Simpli ed Block Diagram 11 24 Theory of Operation ...
Page 249: ...Figure IDC5S11001 here Figure 11 8 4396B Source Group Block Diagram Theory of Operation 11 29 ...
Page 254: ...Figure 12 1 Top View Major Assemblies 12 4 Replaceable Parts ...
Page 290: ...Figure 12 36 Main Frame Assembly Parts 17 19 12 40 Replaceable Parts ...
Page 294: ......
Page 308: ......
Page 311: ...Figure C 1 Power Cable Supplied Power Requirement C 3 ...
Page 312: ......
Page 324: ......