![Agilent Technologies Agilent 4396B Service Manual Download Page 173](http://html.mh-extra.com/html/agilent-technologies/agilent-4396b/agilent-4396b_service-manual_2867991173.webp)
Figure
9-1.
Probe
P
ower
Connector
V
oltages
Inspect
the
T
est
Set
This
procedure
checks
the
operation
of
the
RF
transfer
switch
in
the
85046A/B
test
set.
1.
Turn
the
analyzer
power
o.
2.
Connect
the
test
set
to
the
analyzer
.
3.
Turn
the
analyzer
power
on.
4.
Press
4
PRESET
5 ,
4
Meas
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ANALYZER
TYPE
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NETWORK
ANALYZER
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Refl:
REV
S22(B/R)
.
Then
check
that
the
S22
S12
indicator
LED
lits
.
5.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Refl:
FWD
S11
(A/R)
.
Check
that
the
S11
S21
indicator
LED
lits
.
If
the
LED
operations
are
not
expected,
inspect
the
cable
between
the
analyzer
and
the
test
set.
If
the
cable
seems
good,
verify
the
test
set
in
accordance
with
its
manual.
If
the
LED
operations
are
correct,
continue
with
this
chapter
unless
a
test
set
failure
is
suspected.
T
o
troubleshoot
test
set
failures
,
see
the
test
set
manual.
Inspect
the
Calibration
Kit
Inspect
all
of
the
terminations
(load,
open,
and
short)
for
any
damage
.
If
no
damage
is
found,
perform
the
following
procedure
to
verify
the
short
and
open.
If
any
damage
is
found,
replace
the
termination
with
a
good
one
.
V
erify
Shorts
and
Opens
Substitute
a
known
good
short
and
open
of
the
same
connector
type
as
the
terminations
in
question.
If
the
devices
are
not
from
a
standard
calibration
kit,
see
Modifying
Calibration
Kits
in
the
Function
R
eference
to
use
the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MODIFY
[CAL
KIT]
function.
Set
aside
the
short
and
open
that
could
be
causing
the
problem.
1.
P
erform
an
S11
1-port
calibration
on
a
port
using
the
good
short
and
open.
Then
press
4
F
ormat
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SMITH
CHART
to
view
the
devices
in
Smith
chart
format.
Accessories
T
roubleshooting
9-5
Summary of Contents for Agilent 4396B
Page 10: ......
Page 32: ......
Page 43: ...Figure 2 7 CAL OUT Level Adjustment Location Adjustments and Correction Constants 2 11 ...
Page 46: ...Figure 2 10 Comb Generator Output 2 14 Adjustments and Correction Constants ...
Page 62: ...Figure 2 26 Final Gain Adjustment Location 2 30 Adjustments and Correction Constants ...
Page 76: ...Figure 3 1 Troubleshooting Organization 3 2 T roubleshooting ...
Page 84: ......
Page 90: ...Figure 5 1 Power Supply Lines Simpli ed Block Diagram 5 2 Power Supply T roubleshooting ...
Page 107: ...Figure 5 12 Power Supply Block Diagram 1 Power Supply T roubleshooting 5 19 ...
Page 108: ...Figure 5 13 Power Supply Block Diagram 2 5 20 Power Supply T roubleshooting ...
Page 109: ...Figure 5 14 Power Supply Block Diagram 3 Power Supply T roubleshooting 5 21 ...
Page 110: ......
Page 112: ...Figure 6 1 Digital Control Group Simpli ed Block Diagram 6 2 Digital Control T roubleshooting ...
Page 124: ......
Page 126: ...Figure 7 1 Source Group Block Diagram 7 2 Source Group T roubleshooting ...
Page 160: ...Figure 8 1 Receiver Group Simpli ed Block Diagram 8 2 Receiver Group T roubleshooting ...
Page 168: ......
Page 184: ...Figure 10 6 External Test Setup 1 Figure 10 7 External Test Setup 2 10 10 Service Key Menus ...
Page 185: ...Figure 10 8 External Test Setup 3 Figure 10 9 External Test Setup 4 Service Key Menus 10 11 ...
Page 226: ...Figure 11 3 Power Supply Functional Group Simpli ed Block Diagram 11 6 Theory of Operation ...
Page 231: ...Figure 11 5 Digital Control Group Simpli ed Block Diagram Theory of Operation 11 11 ...
Page 235: ...Figure 11 6 Source Simpli ed Block Diagram Theory of Operation 11 15 ...
Page 244: ...Figure 11 7 Receiver Simpli ed Block Diagram 11 24 Theory of Operation ...
Page 249: ...Figure IDC5S11001 here Figure 11 8 4396B Source Group Block Diagram Theory of Operation 11 29 ...
Page 254: ...Figure 12 1 Top View Major Assemblies 12 4 Replaceable Parts ...
Page 290: ...Figure 12 36 Main Frame Assembly Parts 17 19 12 40 Replaceable Parts ...
Page 294: ......
Page 308: ......
Page 311: ...Figure C 1 Power Cable Supplied Power Requirement C 3 ...
Page 312: ......
Page 324: ......