Getting Started
1
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
29
If you see the waveform, but the square wave is not shaped correctly as
shown above, perform the procedure
“Compensate Passive Probes"
on
page 29.
If you do not see the waveform, make sure the probe is connected
securely to the front panel channel input BNC and to the left side,
Demo 2, Probe Comp terminal.
How AutoScale
Works
Auto Scale analyzes any waveforms present at each channel and at the
external trigger input. This includes the digital channels, if connected.
Auto Scale finds, turns on, and scales any channel with a repetitive
waveform that has a frequency of at least 25 Hz, a duty cycle greater than
0.5%, and an amplitude of at least 10 mV peak- to- peak. Any channels that
do not meet these requirements are turned off.
The trigger source is selected by looking for the first valid waveform
starting with external trigger, then continuing with the lowest number
analog channel up to the highest number analog channel, and finally (if
digital probes are connected) the highest number digital channel.
During Auto Scale, the delay is set to 0.0 seconds, the horizontal time/div
(sweep speed) setting is a function of the input signal (about 2 periods of
the triggered signal on the screen), and the triggering mode is set to Edge.
Compensate Passive Probes
Each oscilloscope passive probe must be compensated to match the input
characteristics of the oscilloscope channel to which it is connected. A
poorly compensated probe can introduce significant measurement errors.
1
Input the Probe Comp signal (see
“Input a Waveform"
on page 27).
2
Press
[Default Setup]
to recall the default oscilloscope setup (see
“Recall
the Default Oscilloscope Setup"
on page 27).
3
Press
[Auto Scale]
to automatically configure the oscilloscope for the
Probe Comp signal (see
“Use Auto Scale"
on page 28).
4
Press the channel key to which the probe is connected (
[1]
,
[2]
, etc.).
5
In the Channel Menu, press
Probe
.
6
In the Channel Probe Menu, press
Probe Check
; then, follow the
instructions on- screen.
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...