Digital Channels
6
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
93
3
Press the
D7 - D0
softkey, then select a logic family preset or select
User
to define your own threshold.
The threshold you set applies to all channels within the selected D7 - D0
group. Each of the two channel groups can be set to a different threshold
if desired.
Values greater than the set threshold are high (1) and values less than the
set threshold are low (0).
If the
Thresholds
softkey is set to
User
, press the
User
softkey for the
channel group, then turn the Entry knob to set the logic threshold. There
is one
User
softkey for each group of channels.
To reposition a digital channel
1
Make sure the multiplexed scale and position knobs to the right of the
key are selected for digital channels.
If the arrow to the left of the
[Digital]
key is not illuminated, press the
key.
2
Use the multiplexed Select knob to select the channel.
The selected waveform is highlighted in red.
3
Use the multiplexed Position knob to move the selected channel
waveform.
If a channel waveform is repositioned over another channel waveform,
the indicator at the left edge of the trace will change from
D
nn
designation (where nn is a one or two digit channel number) to
D
*. The
"*" indicates that two channels are overlaid.
Logic family
Threshold Voltage
TTL
+1.4 V
CMOS
+2.5 V
ECL
–1.3 V
User
Variable from –8 V to +8 V
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...