Measurements
14
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
193
+ Width
+ Width
is the time from the middle threshold of the rising edge to the
middle threshold of the next falling edge. The X cursors show the pulse
being measured. The Y cursor shows the middle threshold point.
– Width
– Width
is the time from the middle threshold of the falling edge to the
middle threshold of the next rising edge. The X cursors show the pulse
being measured. The Y cursor shows the middle threshold point.
Duty Cycle
The duty cycle of a repetitive pulse train is the ratio of the positive pulse
width to the period, expressed as a percentage. The X cursors show the
time period being measured. The Y cursor shows the middle threshold
point.
Rise Time
The rise time of a signal is the time difference between the crossing of the
lower threshold and the crossing of the upper threshold for a
positive- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete rising edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Fall Time
The fall time of a signal is the time difference between the crossing of the
upper threshold and the crossing of the lower threshold for a
negative- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete falling edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Duty cycle =
+ Width
Period
× 100
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...