Save/Recall (Setups, Screens, Data)
18
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
231
•
Setup Info
— when enabled, setup information (vertical, horizontal,
trigger, acquisition, math, and display settings) is also saved in a
separate file with a TXT extension.
•
Length
— sets the number of data points that will be output to the
file. For more information, see
"Length Control"
on page 231.
•
Save Seg
— when data is acquired to segmented memory, you can
specify whether the currently displayed segment is saved or all
acquired segments are saved. (See also
"Saving Data from Segmented
Memory"
on page 168.)
4
Finally, press the
Press to Save
softkey.
A message indicating whether the save was successful is displayed.
See Also
•
"Binary Data (.bin) Format"
on page 289
•
"CSV and ASCII XY files"
on page 296
•
"Minimum and Maximum Values in CSV Files"
on page 297
Length Control
The
Length
control is available when saving data to CSV, ASCII XY, or BIN
format files. It sets the number of data points that will be output to the
file. Only displayed data points are saved.
The maximum number of data points depends on these things:
•
Whether acquisitions are running. When stopped, data comes from the
raw acquisition record. When running, data comes from the smaller
measurement record.
•
Whether the oscilloscope was stopped using
[Stop]
or
[Single]
. Running
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
•
Whether only one channel of a pair is turned on. (Channels 1 and 2 are
one pair, channels 3 and 4 are the other.) Acquisition memory is
divided among the channels in a pair.
•
Whether reference waveforms are on. Displayed reference waveforms
consume acquisition memory.
•
Whether digital channels are on. Displayed digital channels consume
acquisition memory.
•
Whether segmented memory is on. Acquisition memory is divided by
the number of segments.
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...