UART/RS232 Triggering and Serial Decode
•
Rx Stop Bit
— Triggers when a stop bit occurs on Rx. The trigger will
occur on the first stop bit. This is done automatically whether the
device under test uses 1, 1.5, or 2 stop bits. You do not need to
specify the number of stop bits used by the device Under test.
•
Rx Data
— Triggers on a data byte that you specify. For use when the
device under test data words are from 5 to 8 bits in length (no 9th
(alert) bit).
•
Rx 1:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 1. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
•
Rx 0:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 0. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
•
Rx X:Data
— For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers on a data byte
that you specify regardless of the value of the 9th (alert) bit. The
specified data byte applies to the least significant 8 bits (excludes
the 9th (alert) bit).
•
Similar choices are available for Tx.
•
Rx or Tx Parity Error
— Triggers on a parity error based on the parity
you have set in the Bus Configuration Menu.
6
If you choose a trigger condition that includes "
Data
" in its description
(for example:
Rx Data
), then press the
Data is
softkey, and choose an
equality qualifier. You can choose equal to, not equal to, less than, or
greater than a specific data value.
7
Use the
Data
softkey to choose the data value for your trigger
comparison. This works in conjunction with the
Data is
softkey.
8
Optional: The
Burst
softkey lets you trigger on the Nth frame (1- 4096)
after an idle time you select. All trigger conditions must be met for the
trigger to occur.
9
If
Burst
is selected, an idle time (1 µs to 10 s) can be specified so that
the oscilloscope will look for a trigger condition only after the idle time
has past. Press the
Idle
softkey and rotate the Entry knob to set an idle
time.
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...