Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
23
CAN/LIN Triggering and Serial Decode
5
Press
Source
; then, select the channel for the CAN signal.
The label for the CAN source channel is automatically set.
6
Press the
Threshold
softkey; then, turn the Entry knob to select the CAN
signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become the
trigger level when the trigger type is set to the selected serial decode
slot.
7
Press the
Baud
softkey; then, turn the Entry knob to select the baud
rate that matches match your CAN bus signal.
The CAN baud rate can be set to predefined baud rates from 10 kb/s
up to 5 Mb/s or a user- defined baud rate from 10.0 kb/s to 4 Mb/s in
increments of 100 b/s. Fractional user- defined baud rates between
4 Mb/s and 5 Mb/s are not allowed.
The default baud rate is 125 kb/s
If none of the pre- defined selections match your CAN bus signal, select
User Defined
; then, press the
User Baud
softkey and turn the Entry knob
to enter the baud rate.
8
Press the
Sample Point
softkey; then, turn the Entry knob to select the
point between phase segments 1 and 2 where the state of the bus is
measured. This controls the point within the bit's time where the bit
value is captured.
9
Press the
Signal
softkey and select the type and polarity of the CAN
signal. This also automatically sets the channel label for the source
channel.
•
CAN_H
— The actual CAN_H differential bus.
•
Differential (H-L)
— The CAN differential bus signals connected to an
analog source channel using a differential probe. Connect the probe's
positive lead to the dominant- high CAN signal (CAN_H) and connect
the negative lead to the dominant- low CAN signal (CAN_L).
60%
70%
80%
Sample Point
One Bit
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...