Utility Settings
20
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
255
1
Press
[Utility] > Options > Rear Panel
.
2
In the Rear Panel Menu, press
Trig Out
; then, turn the Entry knob to
select from:
•
Triggers
— Each time the oscilloscope triggers, a rising edge occurs on
TRIG OUT. The rising edge is delayed 30 ns from the oscilloscope's
trigger point. The output level is 0- 5 V into an open circuit, and
0- 2.5 V into 50
Ω
. See
Chapter 10
, “Triggers,” starting on page 123.
•
Mask
— The pass/fail status is evaluated periodically. When the
evaluation of the testing period results in a failure, the trigger output
pulses high (+5 V). Otherwise, the trigger output remains at low
(0 V). See
Chapter 15
, “Mask Testing,” starting on page 199.
•
Waveform Generator Sync Pulse
— All of the waveform generator output
functions (except DC and Noise) have an associated Sync signal:
The Sync signal is a TTL positive pulse that occurs when the
waveform rises above zero volts (or the DC offset value).
See
Chapter 17
, “Waveform Generator,” starting on page 215.
The TRIG OUT connector also provides the User Cal signal. See
"To
perform user calibration"
on page 256.
Performing Service Tasks
The Service Menu (under
[Utility] > Service
) lets you perform service- related
tasks:
•
"To perform user calibration"
on page 256
•
"To perform hardware self test"
on page 258
•
"To perform front panel self test"
on page 259
•
"To display oscilloscope information"
on page 259
•
"To display the user calibration status"
on page 259
For other information related to oscilloscope maintenance and service, see:
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...