282
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
22
Reference
•
"Transient Withstand Capability"
on page 283
Oscilloscope Measurement Category
The InfiniiVision oscilloscopes are intended to be used for measurements
in Measurement Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on circuits not
directly connected to MAINS. Examples are measurements on circuits not
derived from MAINS, and specially protected (internal) MAINS derived
circuits. In the latter case, transient stresses are variable; for that reason,
the transient withstand capability of the equipment is made known to the
user.
Measurement category II is for measurements performed on circuits
directly connected to the low voltage installation. Examples are
measurements on household appliances, portable tools and similar
equipment.
Measurement category III is for measurements performed in the building
installation. Examples are measurements on distribution boards,
circuit- breakers, wiring, including cables, bus- bars, junction boxes,
switches, socket- outlets in the fixed installation, and equipment for
industrial use and some other equipment, for example, stationary motors
with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of
the low- voltage installation. Examples are electricity meters and
measurements on primary overcurrent protection devices and ripple
control units.
WA R N I N G
Use this instrument only for measurements within its specified measurement
category.
Summary of Contents for 2000 X-Series
Page 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Page 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Page 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Page 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Page 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Page 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Page 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Page 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Page 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Page 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Page 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Page 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Page 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Page 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Page 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...