4 Maintenance Procedures
158
March 14, 2014, 715004387 Rev. B
15. Carefully push the PEEK union/UNF coupling assembly and capillary
into the probe assembly so that the locating pin on the UNF coupling is
fully engaged in the locating slot at the head of the probe assembly.
16. Fit the nebulizer adjuster knob onto the PEEK union/UNF coupling
assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the gasket and end cover onto the probe assembly.
19. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 screws retaining the probe end-cover.
20. Return the combined 2.5-mm Allen wrench and cone extraction tool to
its storage location on the source adaptor housing.
21. Fit the metal gasket onto the probe tip.
22. Fit the probe tip over the capillary, and screw the tip onto the probe
assembly.
23. Use the 10-mm wrench to tighten the probe tip.
24. Use the nebulizer adjuster knob to adjust the capillary so that it
protrudes by approximately 0.5 mm from the end of the probe tip.
25. Fit the ESI probe into the source (see also
).
Notice:
To avoid gas leakage, fully tighten the probe tip.
UNF coupling locating pin
Probe assembly locating slot
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...