Emptying the exhaust trap bottle
March 14, 2014, 715004387 Rev. B
101
Required material
Chemical-resistant, powder-free gloves
To empty the nitrogen exhaust trap bottle:
1. To stop the LC flow, follow the action below for your software:
2. Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
3. Unscrew and remove the nitrogen exhaust trap bottle from the cap and
associated fittings.
4. Dispose of the waste liquid in accordance with local environmental
regulations.
5. Fit and fully tighten the nitrogen exhaust trap bottle to the cap.
6. Secure the nitrogen exhaust trap bottle in the upright position.
7. Close the source enclosure.
Tip:
A leak test runs automatically. If the test results in a failure, ensure
that the nitrogen exhaust trap bottle is fully tightened to the cap.
Software
Action
MassLynx
In the instrument console, click Stop Flow
.
UNIFI
On the System Console tool bar, click Stop Flow
.
Warning:
To avoid personal contamination with biologically
hazardous, or toxic materials, wear clean,
chemical-resistant, powder-free gloves when handling the
nitrogen exhaust trap bottle. The waste liquid in the bottle
comprises ACQUITY UPLC solvents and analytes.
Warning:
To avoid spreading contamination, dispose of the
waste liquid in accordance with local environmental
regulations.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...