![Waters Xevo TQD Скачать руководство пользователя страница 135](http://html2.mh-extra.com/html/waters/xevo-tqd/xevo-tqd_overview-and-maintenance-manual_3711421135.webp)
Cleaning the ion block assembly
March 14, 2014, 715004387 Rev. B
135
5. Inspect each component for persisting contamination.
Requirement:
If contamination is present, dispose of the component, and
obtain a new one before reassembling the ion block.
Assembling the source ion block assembly
Required materials
•
Clean, chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Needle-nose pliers
•
Isopropyl alcohol in a small container
To assemble the ion block assembly:
1. Fit the extraction cone seal to the ion block.
2. Fit the extraction cone handle’s insulator onto the extraction cone
handle.
3. Fit the extraction cone onto the ion block.
4. Fit the 2 PEEK extraction cone retainer clips onto the ion block.
Warning:
To avoid spreading contamination, dispose of the
components in accordance with local environmental
regulations.
Notice:
• To avoid recontaminating the ion block assembly, wear clean
chemical-resistant, powder-free gloves during this procedure.
• To avoid damaging the fragile extraction cone, do not place it on its
tip; instead, place it on its flanged base.
Notice:
To avoid damaging the extraction cone aperture, take
great care when fitting the extraction cone to the ion block.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...