![Waters Xevo TQD Скачать руководство пользователя страница 185](http://html2.mh-extra.com/html/waters/xevo-tqd/xevo-tqd_overview-and-maintenance-manual_3711421185.webp)
Replacing the roughing pump’s oil demister element
March 14, 2014, 715004387 Rev. B
185
6. Holding the oil demister element slightly elevated to prevent the loss of
the spring, remove the exhaust flange from the oil demister element.
7. Remove the spring from the oil demister element.
8. Dispose of the oil demister element in accordance with local
environmental regulations.
Warning:
To avoid spreading contamination, dispose of the
oil demister element in accordance with local environmental
regulations. It can be contaminated with irritant oil, or with
biohazardous or toxic materials.