A Safety Advisories
204
March 14, 2014, 715004387 Rev. B
Mass spectrometer shock hazard
The following warning applies to all Waters mass spectrometers.
The following warning applies to certain mass spectrometers when they are in
Operate mode.
Mass spectrometer flammable solvents warning
This warning applies to mass spectrometers performing an analysis that
requires the use of flammable solvents.
Warning:
To avoid electric shock, do not remove the mass spectrometer’s
protective panels. The components they cover are not user-serviceable.
Warning:
To avoid nonlethal electric shock, ensure the mass
spectrometer is in Standby mode before you touch any of its external
surfaces that are marked with this high voltage warning symbol.
Warning:
To prevent ignition of flammable solvent vapors in the
enclosed space of a mass spectrometer’s ion source, ensure that nitrogen
flows continuously through the source. The nitrogen supply pressure
must not fall below 400 kPa (4 bar, 58 psi) during an analysis requiring
the use of flammable solvents. Also a gas-fail device must be installed, to
interrupt the flow of LC solvent should the nitrogen supply fail.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...