4 Maintenance Procedures
116
March 14, 2014, 715004387 Rev. B
Assembling the sampling cone assembly
Required material
Clean, chemical-resistant, powder-free gloves
To assemble the sampling cone assembly:
1. Fit the cone gas nozzle handle onto the cone gas nozzle and turn the
handle clockwise to tighten.
2. Carefully fit the sample cone into the cone gas nozzle.
3. Fit the O-ring into the groove created between the sample cone and cone
gas nozzle. (Fit a new O-ring if the old one has been disposed of.)
Notice:
• To avoid recontaminating the sampling cone assembly, wear clean
chemical-resistant, powder-free gloves during this procedure.
• To avoid damaging the fragile sample cone, do not place it on its tip;
instead, place it on its flanged base.
TP02663
O-ring
Sample cone
Cone gas nozzle
Cone gas nozzle handle