4 Maintenance Procedures
168
March 14, 2014, 715004387 Rev. B
2. Gripping the probe heater as shown, carefully pull it off the probe
assembly.
3. Dispose of the probe heater in accordance with local environmental
regulations.
Notice:
To avoid damaging the probe heater’s electrical
connections, do not twist the heater when removing it from the
probe assembly.
Warning:
To avoid spreading contamination, dispose of the
probe heater in accordance with local environmental
regulations. The probe heater can be contaminated with
biohazardous or toxic materials.
Probe heater