Cleaning the sampling cone assembly
March 14, 2014, 715004387 Rev. B
113
4. Remove the O-ring from the sample cone.
5. If the O-ring shows signs of deterioration or damage, dispose of it in
accordance with local environmental regulations.
6. Unscrew and remove the PEEK cone gas nozzle handle from the cone
gas nozzle.
Warning:
To avoid spreading contamination, dispose of the
O-ring or seal in accordance with local environmental
regulations.
Cone gas nozzle
O-ring
Cone gas nozzle handle