Cleaning the sampling cone assembly
March 14, 2014, 715004387 Rev. B
109
Cleaning the sampling cone assembly
You can remove the sampling cone assembly (the sample cone, O-ring, and
cone gas nozzle) for cleaning without venting the instrument.
Removing the sampling cone assembly from the source
Required material
Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Warning:
To avoid electric shock, before beginning this
procedure, prepare the instrument according the procedure on
“Preparing the instrument for operations on or inside its
.
Warning:
To avoid puncture wounds, take great care while
working with the source enclosure open if one or both of these
conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
Warning:
To avoid burn injuries, take great care while working
with the source enclosure open; the source can be hot.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...