Replacing the IonSABRE II probe heater
March 14, 2014, 715004387 Rev. B
167
Replacing the IonSABRE II probe heater
Replace the IonSABRE II probe heater if it fails to heat the probe.
Removing the IonSABRE II probe heater
Required material
Chemical-resistant, powder-free gloves
To remove the IonSABRE II probe heater:
1. Remove the probe from the source (see
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.