![Waters Xevo TQD Скачать руководство пользователя страница 196](http://html2.mh-extra.com/html/waters/xevo-tqd/xevo-tqd_overview-and-maintenance-manual_3711421196.webp)
4 Maintenance Procedures
196
March 14, 2014, 715004387 Rev. B
19. Use the O-ring removal kit to carefully remove the shaft mounting
flange O-ring from the APPI source enclosure side.
20. Dispose of the O-rings in accordance with local environmental
regulations.
Warning:
To avoid spreading contamination, dispose of the
O-rings in accordance with local environmental regulations.
They can be contaminated with irritant oil, or with
biohazardous or toxic materials.
Lamp mounting flange O-ring
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...