4 Maintenance Procedures
108
March 14, 2014, 715004387 Rev. B
Cleaning the source components
Clean the sample cone and cone gas nozzle when these conditions apply:
•
The sample cone and cone gas nozzle are visibly fouled.
•
You dismissed LC and sample-related causes for decreased signal
intensity.
.
If cleaning the sample cone and cone gas nozzles fails to increase signal
sensitivity, also clean the extraction cone. See
If cleaning the extraction cone fails to increase signal sensitivity, clean the ion
block and isolation valve. See
.
If cleaning the ion block and isolation valve fails to increase signal sensitivity,
also clean the ion guide assembly. See
.