1 Specifications and Operating Modes
34
March 14, 2014, 715004387 Rev. B
IntelliStart Fluidics system
Functionality
The IntelliStart Fluidics system is a solvent delivery system built into the
mass spectrometer. It delivers sample or solvent directly to the MS probe in
one of three ways:
•
From the LC column.
•
From two integral reservoirs. Use standard reservoir bottles (15-mL) for
instrument setup and calibration. Use 1.5-mL, low-volume vials (sold
separately) to infuse smaller volumes (see
The reservoirs can also deliver sample through direct or combined
infusion to enable optimization at analytical flow rates.
•
From a wash reservoir that contains solvent for automated flushing of
the instrument’s solvent delivery system.
The onboard system incorporates a selector valve, an infusion pump, and two
sample reservoirs mounted on the bottom, right-hand side of the instrument.
Recommendation:
Use reservoir A for the calibrant solution and tuning
compounds, and reservoir B for analyte/optimization solution.
IntelliStart Fluidics system:
P
W
R
B
A
S
LC
LC
Column
Xevo
TQD probe
Pump
Reservoirs
B
Wash
Waste
Selector valve
A
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...