Replacing the IonSABRE II probe sample capillary
March 14, 2014, 715004387 Rev. B
161
4. Remove the end cover and gasket.
5. Unscrew and remove the nebulizer adjuster knob.
6. Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
Tip:
The PEEK union used with the IonSABRE II probe is notched on
one of its flats, a feature that distinguishes it from the PEEK union used
with the ESI probe (see “Replacing the ESI probe sample capillary” on
End-cover
Nebulizer adjuster knob
Gasket
PEEK union/UNF coupling assembly
Capillary
Locknut
Notch