Replacing the roughing pump’s oil demister element
March 14, 2014, 715004387 Rev. B
183
Replacing the roughing pump’s oil demister element
Replace the roughing pump’s oil demister element annually.
Note:
This procedure is not required for an Edwards oil-free roughing
pump.
Required materials
•
Chemical-resistant, powder-free gloves
•
6-mm Allen wrench
•
10-mm wrench
To remove the roughing pump oil demister element:
1. Vent and shut down the instrument (see the mass spectrometer’s online
Help for details).
2. Allow the roughing pump to cool.
3. Use the 6-mm Allen wrench to remove the 4 bolts securing the exhaust
flange to the roughing pump.
Warning:
To avoid personal contamination with biohazards
or toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when adding or replacing oil. The
roughing pump oil can be irritant, or contaminated with
biohazardous or toxic analyte accumulated during normal
operation.
Warning:
To avoid burn injuries, take great care while
working with the roughing pump; it can be hot.
Exhaust flange
Securing bolt
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...