![Waters Xevo TQD Скачать руководство пользователя страница 115](http://html2.mh-extra.com/html/waters/xevo-tqd/xevo-tqd_overview-and-maintenance-manual_3711421115.webp)
Cleaning the sampling cone assembly
March 14, 2014, 715004387 Rev. B
115
3. Place the vessels in the ultrasonic bath for 30 minutes.
4. If you used formic acid in the cleaning solution, do as follows:
a. Rinse the components by immersing them in separate glass vessels
containing water and then place the vessels in the ultrasonic bath
for 20 minutes.
b. Remove any residual water from the extraction cone by immersing
it in a glass vessel containing methanol and then place the vessel in
the ultrasonic bath for 10 minutes.
5. Carefully remove the components from the vessels, and blow-dry them
with inert, oil-free gas.
6. Inspect each component for persisting contamination. If contamination
is present, do as follows:
a. Use the wash bottle containing 1:1 methanol/water to rinse the
component over the large beaker.
b. Blow-dry the component with inert, oil-free gas.
7. Inspect each component for persisting contamination. If contamination
is present, dispose of the component, and obtain a new one before
reassembling the sampling cone assembly.
Notice:
To avoid recontaminating the components, wear clean,
chemical-resistant, powder-free gloves for the rest of this
procedure.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...